Home > HAMM > Gas Chromatography / Mass Spectroscopy GAS CHROMATOGRAPHY / MASS SPECTROSCOPY Description of Technique ~ Analytical Information ~ Typical Applications ~ Sample Requirements Gas chromatography / mass spectrometry (GC/MS) is the marriage of two analytical methods into a versatile technique for the identification of complex volatile materials. Gas chromatography (GC) effectively separates the different constituents of the sample for subsequent analysis and identification by mass spectrometry (MS). The chromatographic separation relies on the interaction of the sample with a mobile phase and a
stationary phase within the GC instrument column. The sample is carried through the column by the
mobile phase, typically an inert gas. However, the sample is slowed in its travel through the column as
the sample molecules repeatedly adsorb and desorb from the stationary phase in the column. The
affinity of a particular molecule for the stationary phase determines the retention time of that constituent
in the column. The molecules for each component of
the sample will travel through the column at nearly the
same rate and exit (elute) from the column within a
narrow time band that is specific to that component. The typical GC capillary column consists of a smalldiameter tube with a thin film of a high-molecularweight polymer coated on the inside. The polymer is the stationary phase for the chromatographic process. The mobile phase can be any inert gas, but is typically helium. The instrument also includes a heated injection port to vaporize all volatile constituents of the sample and an oven to keep the constituents in gas form as they pass through the column. As a sample constituent elutes from the GC column, it enters the ionization chamber of the mass spectrometer where the molecules are ionized, typically by electron impact. When an electron impact with a sample molecule results in the loss of an electron from the molecule, a positive ion is formed. Some of the molecular ions are further fragmented into daughter ions and neutral fragments. The
positive ions are then repelled out of the ionization chamber by a small positive charge within the
chamber. Negative ions are also formed by the electron impact, but are not analyzed.
The positive ions are separated according to their mass by a mass analyzer. The mass analyzer most
commonly used in GC/MS is the quadrapole filter, in which the ions pass by four hyperbolic magnetic
poles created by a radio frequency field. The magnetic poles separate the ions by their mass/charge
ratio, successively focusing ions with increasing mass onto a detector for counting. The analyzer scans
step-wise through a set range of mass values to evaluate the relative abundance of ions at each mass
value. The quadrapole filter can perform a complete mass scan within the duration of a single GC
elution band. Material Identification - The first result from the compiled data is a total-ion chromatogram (TIC), which is a plot of the total mass eluting from the GC and detected by MS as a function of time. Each peak or band in the chromatogram represents a discrete chemical compound, or a mixture of compounds with identical retention times. The retention times in the chromatogram provide the first indication of the sample constituents. More specific identification of the compound(s)
for each band can then be made from the mass
spectrum corresponding to the band. Compounds
are identified from the mass spectrum by
their unique ion fragmentation patterns. This
compound identification analysis is performed by
a computerized comparison of the mass spectra
for the sample with spectra library for known Quantitation - The analysis results can be quantified using the data from the chromatogram. The area under each peak in the chromatogram is proportional to the concentration of the ompounds represented by that peak. The concentration for each compound in the sample is calculated from a standard curve of known concentrations established for that compound. The analysis sensitivity can be as low as a few nanograms.